| Linear Peristaltic Volumetric Infusion Pump Keep Vein Open (KVO) Rate: 5 mL/hr or programmed rate, whichever is less Nurse Call: Available through the COMMUNICATIONS PORT Battery: 12 Volt, 2.0 Ah sealed lead-acid Battery Life: Approximately 6 hours with pump running at a rate from 1 to 1400 mL/hr Battery Recharge: 8 hours for more than 80% recharge AC Power Requirements: 110/120V, 60 Hz Power Cord: 2.9 m (9 ft) long, with hospital grade plug Fuse: 0.5 A, 250V, SB, 5.2 mm (13/64 in) x 20 mm (25/32 in) Leakage Current: Less than 50 microamps (using UL-544 specified test methods) Weight: Approximately 5.3 kg (11.6 lbs) Dimensions: 20 cm W x 13 cm D x 29 cm H (7.9" W x 5.1" D x 11.4" H) Flow Rate Range: Primary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-1999 mL/hr in 1 mL increments. Upper limit can be reduced by authorized service personnel |